Operating Modes |
Eddy Current Modes |
C-Scan |
Eddy Current Single Probe Frequency |
Eddy Current Dual Probe Frequency |
Eddy Current 2 Probe, 2 Channel |
Rotating Drive |
Conductivity with thickness |
Probe |
Connectors |
Simultaneous probe operation possible using LEMO 12 way and LEMO 4 way |
12-Way LEMO 2B (Absolute, Bridge, Reflection, Rotary, Conductivity) |
4-Way LEMO 0B (Bridge, Reflection) |
Rotary |
600-3000 rpm. ETher Mercury Drive (ADR002) and Saturn (ARD001), Hocking 33A100, Rohmann MR3, SR1 and SR2 Drive (special adapter needed) |
Conductivity |
Option becomes active with use of AMCheck conductivity probe and cable. |
Frequency |
Dual Frequency |
Frequency Dual Frequency 10Hz - 12.8MHz & Mix -18 to +18dB on output |
Gain |
Overall |
-18 to + 104dB, 0.1, 1 and 6dB steps (104dB maximum) |
Input |
0dB or 12dB |
Drive |
0dB, 6dB and 10dB (0dB reference 1mW into 50 ohm). |
Max X/Y Ratio |
+/-100.0 dB |
Phase |
Range |
0.0-359.9°, 0.1° steps |
Filters |
High Pass |
DC to 2kHz or Low Pass Filter, which ever is lower in 1Hz steps. Plus variable adaptive balance drift compensation 0.01 - 0.5 Hz (6 steps). |
Low Pass |
1Hz to 2kHz or a quarter of the lowest test frequency, which ever is lower in 1 Hz steps. |
Balance Load |
Manual |
14 internal balance loads; 2.2μH, 5.0μH, 6.0μH, 6.5μH, 7.0μH, 7.5μH, 8.2μH, 12μH, 15μH, 18μH, 22μH, 30μH, 47μH, 82μH |
Automatic |
Optimized balance load selection. |
Mix Channel |
Frequency |
Full frequency range available on both channels |
Probe Mode |
Simultaneous reflection / bridge and absolute including simultaneous two probe Differential and Absolute |
Mix Gain |
X/Y -18 to +18dB |
Mix Phase |
0.0-359.9°, 0.1° steps |
Alarm Gates |
Box |
Fully configurable, Freeze, Tone or Visual. |
Sector |
Fully configurable, Freeze, Tone or Visual. |
Display |
Type |
5.7” (145mm), 18 bit Color, daylight readable. |
Viewable Area |
115.2mm (Horizontal) x 86.4mm (Vertical) |
Resolution |
640 x 480 pixels |
Flip |
Manual or automatic screen orientation change to enable left or right handed use. |
Configurable Screen |
Full Screen, Single, Dual Spot or Dual Pane with variable size and location and function e.g. XY, Timebase, Waterfall and Meter. |
Color Schemes |
User configurable Dark, Bright and Black & White |
Display Modes |
Spot, Time base (0.1-20 seconds x 1-200 sweeps, up to 55 seconds), Waterfall, Meter with peak hold and % readout, Distance (single axis, changes with direction), Strip Chart (single axis, unidirectional) and C-Scan. |
Graticules |
None, Grid (4 sizes 5, 10, 15 and 20% FSH), Polar (4 sizes 5, 10, 15 and 20% FSH) |
Offset |
Spot Position: Y =-50 to +50, X =-65 to +65% |
Digital Spot Position Readout |
Display in X,Y or R,θ |
Summary |
Display of all settings in Legacy Format |
Removable Data Storage |
Media |
Micro SD HC Card 32GB |
Setup Storage |
Over 10,000 settings |
Stored Screen Shots |
micro SD up to 32GB, holding over 10,000 screen shots |
Recorded Data |
Over 500 2.5 minute long data recordings. |
Guides |
10,000 Slides plus |
C-Scan |
Max no of C-Scan Data Files 1,000 |
Advanced Features |
Data Logging |
Real-time recording of signal data and Replay on instruments and desktop PC up to 164 seconds |
Guides |
Create and display a slide show containing instructions, tutorials and procedures using Microsoft PowerPoint. |
Attachments |
Screenshots and Data Recordings are saved in a folder with the name of the Settings. |
Loop |
Capture a live repetitive signal and then optimize the instrument settings (Phase, Gain, Filters) to simplify optimizing the parameters |
Trace |
Allows a calibration reference signal to be stored on the screen and then compared with the live signal |
Auto Phase |
Allows phase angle to be automatically set to a pre-set angle |
Scanning |
Connector |
8 way LEMO 1b for encoder and scanner control |
Encoder |
2 phase 2 axis; =X/Y or R-Theta |
Automatic |
Controls and acquires data from a Stepper Motor Driven XY Scanner |
Count Rate Max |
100kHz |
C-Scan |
Resolution |
Max size 1 million data points |
Scaling |
0.1-999.9 pixels/mm. |
Typical Scan |
120 by 100 mm at 0.1mm resolution. |
Data Saved |
Data stored as XY Pairs for 2 Channels. Data presentation X, Y, R or theta on CH1, Ch2 or Mix. |
Outputs |
PC Connectivity |
Open collector transistor (32v dc at 10mA max) available on 12 way LEMO. |
Digital Volt Free Alarm |
On Lemo 12 way Open collector transistor (36v dc at 10mA max). |
VGA |
Full 15 way VGA output (EC screens only) |
Languages |
Selectable from English, French, Spanish, Italian, Portuguese, Russian, Japanese, Chinese, Turkish, Czech, and Norwegian. |
Verification Levels |
The system includes on delivery a 2 year validity Verification Level 2 detailed functional check and calibration as per ISO 15548-1:2013Power-on self test The system performs a self |
Power-on Self Test |
The system performs a self test on start up of external ram, sd ram, accelerometer, Micro SD card, LCDscreen buffer. |
Power |
External |
100-240 v 50-60Hz 30 Watts |
Battery |
Internal 7.2V nominal @ 3100mAh = 22.32 watt.hr |
Running Time |
Up to 8 hours with a 2MHz Pencil Probe 30% Back Light and up to 6 hours with a Rotary Drive 50% duty cycle. |
Charging Time |
2.5 hrs. charge time, Simultaneous charge and operation |
Weight Including Internal Battery |
1.3 kg, 2.9 lbs. |
Physical |
Size (L x W x H) |
5.7" x 2.1 x 9.3" (146 x 53 x 237mm) |
Material |
Aluminum alloy Mg Si 0.5 powder-coated epoxy |
Operating Temperature |
-4° to 140° F (-20 to +60 °C) |
Storage Temperature |
Storage for up to 12 months -4° to 95° F (-20 to +35 °C) Nominal 68° F (+20 °C) |
IP Rating |
IP54 |