Test Range: |
0.4" to 394" (1mm to 10,000mm) |
Velocity: |
1,000 - 10,000 m/s continuously variable |
Probe Zero: |
0 to 1,000 μs |
Delay: |
400" (0 - 10,000m) in steel at 5930m/s (19455f/s) |
Gain: |
0 to 110dB adjustable in 0.5, 1, 2, 6, 10, 14 and 20 dB steps |
Test Modes: |
Pulse Echo and transmit/receive |
Pulser: |
200V fixed. 50nS square wave. Rise/Fall times <15nS into 50R load. |
Damping: |
50 and 400 ohm damping selectable |
P.R.F.: |
Adjustable 5Hz to 1kHz. External sync also available. |
Screen Update Rate: |
60Hz |
Rectification: |
RF, Fill wave, +ve half-wave and -ve half-wave. |
Frequency Range: |
1.0MHz to 14Mhz (-3dB) |
System Linearity: |
Vertical = 0.5% Full Screen Height (FSH). Horizontal = +/- 0.2% Trace Full Screen Width (FSW). |
Reject (Selectable): |
Up to 80% Linear reject (removes baseline noise without affecting indication amplitude) or Up to 50% Suppressive reject (increase zero offset and reduces amplitude of all echoes). LED Warning when active. |
Units: |
Imperial (in), Metric (mm) |
Display: |
Color Transflective VGA (640 x 480) TFT Display Area: 4.57" x 3.43" (116.16 x 87.2mm) A-Scan Area: 400 x 510 pixels (normal), 460 x 620 (FS) Colors: 9 color palette with variable brightness |
Gate Monitor: |
Two independent gates for measurement and monitoring. Start and width fully adjustable over the entire range of the instrument. Levels adjustable from 0% to 100%, positive or negative triggering on each gate with audible & visual alarms. Gate resolution is 5nS. |
Zoom: |
Expands range and delay to cover the area set by Gate 1 start & width controls. |
AGC: |
Automatic Gain Control automatically sets the signal in Gate 1 to a level between 10% and 90% FSH, tolerance between 5% and 20%. |
Measurement Modes: |
Mode 1: |
Signal Monitor, Gate Alarms can be active but no measurements are displayed. |
Mode 2: |
Depth and amplitude of first signal in gate. |
Mode 3: |
Echo-Echo distance measurements. |
Mode 4: |
Trigonometric display of beam-path, surface distance (including X-offset) and depth of indication from the inspection surface together with echo amplitude. Curved surface correction can be applied for convex and concave surfaces. Half-skip can be indicated on screen. |
Mode 5: |
Gate to Gate distance measurement |
Mode 6: |
Flank to Flank |
Mode 7: |
Beam Angle, calculated from beampath, hole radius and hole center depth |
Measurement Display: |
Live display and updates on screen at 10 times per second. Large display of a single measurement available. |
Waveform Smoothing Select from: |
- None (both min and max values are displayed in the A-Scan)
- Fill (Min values set to baseline value, produces a solid A-Scan)
- Smooth (min values ignored, produces a clear outline)
|
Auto-Cal: |
Provides automatic calculation of velocity and probe zero from 2 reference echoes. Automatic calibration with two echoes. |
Reference Waveform: |
Displays a previously stored A-log in a color different from the active display: enabling a quick visual check of the differences. |
Clock: |
Built-in battery backed RTC keeps time and date. Visible on the status line, always stored with Panels, A-logs, etc. |
Internal Memory: |
4GB storage available for A-Scans, panels, T-logs, B-logs, etc. 450,000 Panels, 200,000 A-Logs, 300,000 B-Charts, 440,000 T-Logs. |
Active Peak Memory: |
Retains all A-Scans on screen for echo-dynamic pattern analysis, with the active A-Scan displayed in a separate color. |
Notes: |
Alphanumeric labelling for panel stores, A-logs, B-logs, etc. |
Display Freeze: |
Hold the current waveform on screen for off-line processing. |
Help Key: |
Shows software and hardware information. |
Language Support: |
Multiple languages are selectable from a list including: English, French, Spanish, Russian, Chinese (Modern). Others are available on request. |
Encoder Connection: |
Lemo min 4-pin connector (D50). |
Video Output: |
Factory Option on D-50. |
USB Connection: |
Internal storage shown as Memory Device. |
Transducer Sockets: |
BNC or Lemo (factory option) |
Power: |
Lithium Ion 14.4V battery pack. Typically 12 hours for D-50. Indication of battery charge status. Recharge time 3-4 hrs. Battery can be charged separately. Mains pack optional. |
Charger: |
100-240 VAC, 50-60 Hz |
Environmental: |
Case sealed to IP67 |
Temperature: |
Operating: 14° to 131°F (-10° to 55°C) Storage: -40° to 167°F (-40° to 75°C)
|
Size: |
6.77" x 9.37" x 2.75" (172 x 238 x 70mm) (HxWxD) |
Weight: |
3.7 lbs. (1.7 kg) with battery. |
Warranty: |
2 year |
Extended Warranty: |
Sonacover - extended 5 year warranty, including 4 calibrations |
Calibration Standard: |
EN12668-1:2020 (Detailed Specification available upon request) |
Standards: |
Vibration to 514.5-5 Proc 1 Annex C Fig 6 Shock 516.5 Proc 1 15g/6ms Explosive atmospheres - MIL-STD 810G Methos 511.5, Procedure 1 |