Centurion NDT FM-140XL Digital Conductivity Meter

Centurion NDT FM-140XL Digital Conductivity Meter

Note: This product has been discontinued, the replacement is:

ETher NDE SigmaCheck 2

Centurion NDT FM-140XL Digital Conductivity Meter

Features:

The Centurion NDT Model FM-140XL Conductivity Meter makes conductivity testing simple and worry free for most non-ferrous metal testing. Portable, yet ruggedly designed, it instantly generates readings with a bright, direct digital display, in percent IACS (International Annealed Copper Standard). To prolong battery life, readings only display when probe and testing surfaces meet.

The FM-140XL delivers precision accuracy to +/-.5% of the readings at 20° C. Included is a RS-232 port, which allows the user to download information to computers, micro-controllers or other peripherals. The FM-140XL is housed in a compact, lightweight and shock resistant case, for long life and durability. Placing the probe against the testing surface generates a direct conductivity reading in large brilliant LED lights.

Applications:

The FM-140XL Conductivity Meter is useful for sorting a wide variety of non-ferrous metals and alloys, as well as for measuring changes in physical properties reflecting conductivity. Such changes include hardening, annealing, variations in alloy, heat treating corrosion and/or aging.

Its rechargeable Ni-Cad batteries provide up to 10 hours of continuous operation from one charge. A lift-off compensation circuit easily handles rough surfaces and compensates for probe wobble. Two test samples are mounted atop of the unit for easy set up.

Specifications:

Accuracy: +/- 0.5% at 20°C.
RS-232: Downloads data into the user's computer, micro-processor or other peripheral.
Resolution: 0.1/IACS for any reading.
Readout: 4.0 digits, 0.43" (1.09cm) seven segment LED display with fixed decimal point. Display automatically shuts off whent he probe is lifted from the part being tested.
Lift-off: Automatic lift-off compensation circuit compensates for moderately rough surfaces and minor probe wobble.
Calibration: Two case mounted set-up samples. Separate front panel control for each sample.
Batteries: Two rechargeable, self-contained Ni-Cad battery packs are standard. Approximately 10 hours of continuous operation on a single charge.
Line Operation: May be operated on 115V 50/60 Hz line or 230V 50/60 Hz.
Weight: 5.0 lbs. (2.27 kg).
Dimensions: 2.75" high x 8.5" wide x 9" deep (7cm x 21.6cm x 22.9cm).
Probe: 7/16" diameter, shielded. probe replacement requires calibration at the factory.
Handle: Multi-adjustable handle enables operator to lock unit into position for easiest viewing angle.


Product Literature

Description Part # Saved Price Select Option Add to
Centurion NDT FM-140XL Digital Conductivity Meter EC-CNDT-220220A Request Pricing